Vis enkel innførsel

dc.contributor.authorKnudsen, Hans Petter
dc.date.accessioned2009-01-08T12:10:03Z
dc.date.issued1985
dc.identifier.urihttp://hdl.handle.net/11250/114504
dc.description.abstractHigh precision in hydroacoustic instruments for fisheries research demands high precision and reliability in the test instruments. To simplify the test measurements, a special test unit has been developed. The operating principles are described, and a technical description is given for this special-purpose signal generator. Application of the unit is described in detail, and an easy way to verify the accuracy of signal amplitude and frequency is recommended. Finally a computational example to derive correction factors is given.en
dc.format.extent544195 bytes
dc.format.mimetypeapplication/pdf
dc.language.isoengen
dc.publisher[Fiskeridirektoratets havforskningsinstitutt]en
dc.relation.ispartofseriesFiskeridirektoratets skrifter, Serie Havundersøkelseren
dc.relation.ispartofseriesvol 17 no 16en
dc.titleTime-Amplitude-Frequency (T-A-F) - A special electronic unit for measuring the TVG function of research echo soundersen
dc.typeResearch reporten
dc.source.pagenumbers. 529-541en


Tilhørende fil(er)

Thumbnail

Denne innførselen finnes i følgende samling(er)

Vis enkel innførsel