dc.contributor.author | Foote, Kenneth G. | |
dc.date.accessioned | 2012-08-15T14:42:29Z | |
dc.date.available | 2012-08-15T14:42:29Z | |
dc.date.issued | 1989 | |
dc.identifier.citation | This report is not to be cited without prior reference to the author | no_NO |
dc.identifier.uri | http://hdl.handle.net/11250/104565 | |
dc.description.abstract | Calibration of scientific echo sounders and integrators entails,
inter alia, measurement of the dynamic range. This can be done by
repeating on-axis measurements of echoes from each of two different
standard targets. When a large dynamic range is to be tested, use of a
reflector may be convenient. An example is given for which a target
strength of roughly 15 dB was desired for use at 38 kHz. The solution:
a 12-mm-thick copper disk of 50-cm diameter, is described. Results from
its first field applications, where linear operation of a new acoustic
system over a 56-dB range was established, are given. | no_NO |
dc.language.iso | eng | no_NO |
dc.publisher | ICES | no_NO |
dc.relation.ispartofseries | ICES CM Documents;1989/B:4 | |
dc.subject | echo sounder | no_NO |
dc.subject | ekkolodd | no_NO |
dc.subject | gear experiments | no_NO |
dc.subject | redskapsforsøk | no_NO |
dc.title | Calibration reflector | no_NO |
dc.type | Working paper | no_NO |
dc.subject.nsi | VDP::Agriculture and fishery disciplines: 900::Fisheries science: 920::Fisheries technology: 924 | no_NO |
dc.source.pagenumber | 10 s. | no_NO |