dc.contributor.authorGytre, Trygve
dc.date.accessioned2012-08-02T13:35:34Z
dc.date.available2012-08-02T13:35:34Z
dc.date.issued1988
dc.identifier.citationThis report is not to be cited without prior reference to the authorno_NO
dc.identifier.urihttp://hdl.handle.net/11250/104434
dc.description.abstractThe paper describes a system for simplified calibration of a new STD instrument. When measuring, the raw data from the instrument conductivity, temperature and pressure sensors are fed into a microprocessor inside the instrument. The microprocessor computes the calibrated primary data by inserting the raw data into an equation of the form Y= A+B.X+C.Xexp2 where A,B and C are calibration coefficients. The new instrument automatically computes and inserts these calibration coefficients itself when exposed to a temperature bath, a conductivity bath and a pressure reference with a few known values.no_NO
dc.language.isoengno_NO
dc.publisherICESno_NO
dc.relation.ispartofseriesICES CM Documents;1988/C:13
dc.subjectgear experimentsno_NO
dc.subjectredskapsforsøkno_NO
dc.titleAutomatic calibration of the sensors used in a miniature STD instrumentno_NO
dc.typeWorking paperno_NO
dc.subject.nsiVDP::Agriculture and fishery disciplines: 900::Fisheries science: 920::Fisheries technology: 924no_NO
dc.source.pagenumber6 s.no_NO